RT Journal Article T1 Lenstar LS900 vs EchoScan US-800: comparison between optical and ultrasound biometry with and without contact lenses and its relationship with other biometric parameters A1 Noya Padín, Verónica A1 García Queiruga, Jacobo A1 Iacubitchii, Maria A1 Giráldez Fernández, María Jesús A1 Yebra-Pimentel Vilar, Eva A1 Pena Verdeal, Hugo K1 Ocular biometry K1 Lenstar LS900 K1 EchoScan US800 K1 Contact lenses K1 Human morphology AB Background: Due to the increasing use of contact lenses (CL) and the interest in ocular and body size relationships, this study aimed to compare measurements from two biometers (contact ultrasonic EchoScan US-800 and non-contact optical Lenstar LS900) with and without CL and to explore the relationship between ocular and body biometric parameters.Design and Methods: This cross-sectional study measured ocular biometry using two biometers along with their body height and right foot length in 50 participants. Differences between biometry data from the two devices were compared and correlations between ocular and body biometric values were analyzed.Results: All parameters showed interbiometric differences (p≤0.030), except crystalline lens thickness during CL wear (p=0.159). Comparing measurements with and without CL, differences were observed in axial length (p<0.001), vitreous length measured by optical biometer (p=0.016), and anterior chamber depth by ultrasonic biometer (p<0.016). Lens thickness remained unaffected (p≥0.190). Body height and foot length were correlated with anterior chamber depth, vitreous length, and axial length (p≤0.019, r≥0.330). Most biometric parameters were correlated among them using both devices (p≤0.037, r≥0.296).Conclusions: These biometers are not interchangeable and CL affects measurements. Body height and foot length correlate with ocular dimensions, and most ocular biometric values correlate positively. PB Taylor and Francis SN 1745-2422 YR 2023 FD 2023-07-07 LK http://hdl.handle.net/10347/33545 UL http://hdl.handle.net/10347/33545 LA eng NO Noya-Padin V, Garcia-Queiruga J, Iacubitchii M, Giraldez MJ, Yebra-Pimentel E, Pena-Verdeal H. Lenstar LS900 vs EchoScan US-800: comparison between optical and ultrasound biometry with and without contact lenses and its relationship with other biometric parameters. Expert Rev Med Devices. 2023;20(8):681-690 NO This is an Accepted Manuscript of an article published by Taylor & Francis in Expert Review of Medical Devices on 7 Jul 2023, available at: https://doi.org/10.1080/17434440.2023.2233410 DS Minerva RD 22 abr 2026