Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers

dc.contributor.affiliationUniversidade de Santiago de Compostela. Facultade de Óptica e Optometríaes_ES
dc.contributor.authorNoya Padín, Verónica
dc.contributor.authorPena Verdeal, Hugo
dc.contributor.authorGarcía Queiruga, Jacobo
dc.contributor.authorNores Palmás, Noelia
dc.contributor.authorCastro Giráldez, Alba
dc.contributor.authorYebra-Pimentel Vilar, Eva
dc.date.accessioned2024-07-30T09:08:02Z
dc.date.available2024-07-30T09:08:02Z
dc.date.issued2024
dc.description.abstractBackground: Precise quantification of corneal parameters enhances the clinical of the anterior corneal surface for various clinical assessments, with ongoing research focusing on optimizing everyday practice through device comparison studies. The study aimed to compare the agreement on the anterior corneal surface characteristics values measured by four commercially available devices: EyeSys 2000, Oculus Easygraph, ARK700A and ARK510A. Methods: Observational cross-sectional study involving 50 eyes from healthy volunteer recruited among subjects attending for an eye examination. Corneal radius minimum keratometry reading (min-K) and maximum keratometry reading (max-K), astigmatism axis and eccentricity were measured in all participants with EyeSys, Oculus Easygraph, ARK700A and ARK510A. Measurements were performed three times only in the right eye in random order by four optometrists who were unaware of the results of the other researchers. Results: A statistically significant difference occurred for the measurements of the corneal radius (Greenhouse-Geisser and Huynh-Feldt: P≤0.001 for both max-K and min-K radius) and the axis of astigmatism (Friedman: P<0.001), between the four devices. Pairwise comparisons revealed no radius differences between Oculus Easygraph and ARK510A (Sidak: min-K radius, P=0.231; max-K radius, P=0.224), and no axis differences in all pairwise analyses that have not imply the EyeSys (Wilcoxon: all P≥0.130). There was no statistical difference in the eccentricity value obtained by the topographers (paired t-test: P=0.507). Conclusions: The present study showed that EyeSys 2000 measurements differed from those of the other devices across all parameters. Alternatively, both topographers showed a perfect agreement concerning eccentricity measurements in everyday practice.es_ES
dc.description.peerreviewedSIes_ES
dc.identifier.citationNoya-Padin V, Pena-Verdeal H, García- Queiruga J, Nores-Palmas N, Castro-Giraldez A, Yebra- Pimentel E. Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers. AME Med J 2024es_ES
dc.identifier.doi10.21037/amj-23-257
dc.identifier.issn2520-0518
dc.identifier.urihttp://hdl.handle.net/10347/34551
dc.language.isoenges_ES
dc.publisherAMJes_ES
dc.rightsCREATIVE COMMONS ATTRIBUTION-NONCOMMERCIAL-NODERIVATIVES 4.0 INTERNATIONALes_ES
dc.rights.accessRightsopen accesses_ES
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subjectAgreement corneal topographes_ES
dc.subjectAgreement autokeratometeres_ES
dc.subjectEyeSyses_ES
dc.subjectEasygraphes_ES
dc.subjectARKes_ES
dc.subject.classification220915 Optometríaes_ES
dc.titleComparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometerses_ES
dc.typejournal articlees_ES
dc.type.hasVersionVoRes_ES
dspace.entity.typePublication
relation.isAuthorOfPublicationa33f1b46-c5e2-417e-99e1-317d1e6e3ff6
relation.isAuthorOfPublication8743c628-62df-4f90-a166-4946245a6e1f
relation.isAuthorOfPublication45eab007-782a-4666-aac6-8c7020f1c661
relation.isAuthorOfPublication.latestForDiscoverya33f1b46-c5e2-417e-99e1-317d1e6e3ff6

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
AMJ-23-257-final.pdf
Size:
308.47 KB
Format:
Adobe Portable Document Format
Description:

Collections