Characterization of the Charge Collection Efficiency in Silicon 3-D-Detectors for Microdosimetry
Loading...
Identifiers
Publication date
Advisors
Tutors
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
IEEE
Abstract
New silicon 3-D-microdetectors have been developed to perform microdosimetry measurements for applications in hadron therapy. In this work, the charge collection efficiency (CCE) of an improved second generation of microdetectors having two different thicknesses (10 and 20 \mu \text{m} ) and a diameter of 25 \mu \text{m} has been studied by means of the ion beam induced charge (IBIC) technique. New methods to study the active volume and the CCE of microdosimeters are proposed and verified here. The results show that, for this new generation of microdetectors, the CCE is 100% for radial distances up to 10.2 \mu \text{m} from the center of the device, and it rapidly decays between 10.2 \mu \text{m} and the detector edge. The characterization of the CCE conducted here will allow us to completely explain the energy spectra obtained during the microdosimetry studies performed in clinical centers.
Description
Bibliographic citation
IEEE Transactions on Instrumentation and MeasurementVolume 702021 Article number 9339896
Relation
Has part
Has version
Is based on
Is part of
Is referenced by
Is version of
Requires
Sponsors
This work was supported by the CNRS-Momentum fellow and the European
Union’s Horizon 2020 Research and Innovation program under the Marie
Sklodowska-Curie Grant Agreement no. 745109 and the G. A. no. 654168
Rights
Attribution-NonCommercial-NoDerivatives 4.0 Internacional