Study of scintillation properties and performance of CsI(Tl) detectors over time
Loading...
Identifiers
Publication date
Advisors
Tutors
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
Abstract
This work presents a systematic study of the properties (Light Output Non-Uniformity (LONU) and energy resolution) of two CsI(Tl) scintillation units over a span of almost three years, under adverse conditions of humidity and temperature. These two crystals are part of the CALIFA detector, a highly segmented calorimeter and spectrometer for
rays and light-charged particles, that is placed surrounding the reaction target at Cave C, the experimental cave of the R
B (Reactions with Radioactive Relativistic Beams) collaboration at the GSI-FAIR facilities in Darmstadt, Germany. The findings obtained after the experiment indicate that there was no significant impact on the performance in terms of resolution and LONU. This suggests that the employed wrapping for light collection effectively serves as an excellent barrier against humidity. As a result, the crystal is preserved in a far better condition than initially anticipated.
Description
Bibliographic citation
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Volume 1059, 2024, 169003
Relation
Has part
Has version
Is based on
Is part of
Is referenced by
Is version of
Requires
Sponsors
The first author’s work was partially supported by the Spanish Ministry of Innovation and Science (grant PRE2019-087415).
Rights
Atribución-NoComercial 4.0 Internacional
© 2023 The Authors. Published by Elsevier B.V. This is an open access article under the CC BY-NC license
© 2023 The Authors. Published by Elsevier B.V. This is an open access article under the CC BY-NC license