Abstract
Uncertainties of refractive and group index in dispersion measurement by spectrally resolved white light interferometry are deeply analyzed. First, the contribution to uncertainty of the different parameters affecting both indices is identified. Afterwards, results are presented for a 1.5 mm thick fused silica sample over a broad spectral range, from 400 to 1000 nm, and the effects that mostly deteriorate the measurement accuracy are established. Finally, the different contributions are quadratically combined to determine the total uncertainty of the two indices
Description
Keywords
Bibliographic citation
Measurement 187 (2022) 110225
Relation
Has part
Has version
Is based on
Is part of
Is referenced by
Is version of
Requires
Publisher version
https://doi.org/10.1016/j.measurement.2021.110225Sponsors
The financial support of the Spanish Ministry of Economy and Competitiveness through the coordinated grants MAT2017-89239-C2-1-P is gratefully acknowledged. Moreover, this work was funded by the Xunta de Galicia and FEDER (ED431D 2017/06, ED431E2018/08, GRC 508 ED431C 2020/10)