Comparison of Central Corneal Thickness measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomography

dc.contributor.affiliationUniversidade de Santiago de Compostela. Departamento de Física Aplicadaes_ES
dc.contributor.authorGonzález Pérez, Javier
dc.contributor.authorQueiruga Piñeiro, Juan
dc.contributor.authorSánchez García, Ángel
dc.contributor.authorGonzález Méijome, José Manuel
dc.date.accessioned2024-02-09T09:31:06Z
dc.date.available2024-02-09T09:31:06Z
dc.date.issued2018
dc.descriptionThis is an Accepted Manuscript version of the following article, accepted for publication in Current Eye Research. [González-Pérez J, Queiruga Piñeiro J, Sánchez García Á, González Méijome JM. Comparison of Central Corneal Thickness Measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomography. Curr Eye Res. 2018 Jul;43(7):866-872]. It is deposited under the terms of the Creative Commons Attribution-NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited.es_ES
dc.description.abstractPurpose: To compare central corneal thickness (CCT) measured by standard ultrasound pachymetry (USP), and three non-contact devices in healthy eyes. Methods: A cross-sectional study of CCT measurement in 52 eyes of 52 healthy volunteers was done by a single examiner at Ocular Surface and Contact Lens Laboratory. Three consecutive measurements were done by standard USP, non-contact tono-pachymeter, Pentacam corneal topographer, and Anterior Segment Optical Coherence Tomography (AS-OCT). The mean values were used for assessment. The results were compared using multivariate ANOVA, linear regression, and Pearson correlation. Agreement among the devices was analyzed using mean differences and Bland-Altman analysis with 95% limits of agreement (LoA). Finally, reliability was analyzed using intraclass correlation coefficient (ICC). Results: Mean CCT by ultrasound pachymeter, tono-pachymeter, corneal topographer and AS-OCT were 558.9 ± 31.2 µm, 525.8 ± 43.1 µm, 550.4 ± 30.5 µm, and 545.9 ± 30.5 µm respectively. There was a significant positive correlation between AS-OCT and USP (Pearson correlation = 0.957, p < 0.001), corneal topography and USP (Pearson correlation = 0.965, p < 0.001), and corneal topography and AS-OCT (Pearson correlation = 0.965, p < 0.001). There was a lower correlation between CT-1P tono-pachymeter and the other three modalities. Intraclass correlation coefficients show an excellent reliability between pairs except for CT-1P against the other three instruments that were found moderate. Conclusions: CT-1P tono-pachymeter underestimates CCT measurements compared to Scheimpflug system, AS-OCT device, and USP. Mean CCT among USP, Pentacam and AS-OCT were comparable and had significant linear correlations. In clinical practice, these three modalities could be interchangeable in healthy patients.es_ES
dc.description.peerreviewedSIes_ES
dc.identifier.citationGonzález-Pérez J, Queiruga Piñeiro J, Sánchez García Á, González Méijome JM. Comparison of Central Corneal Thickness Measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomography. Curr Eye Res. 2018 Jul;43(7):866-872. doi: 10.1080/02713683.2018.1461910. Epub 2018 Apr 13.es_ES
dc.identifier.doi10.1080/02713683.2018.1461910
dc.identifier.essn1460-2202
dc.identifier.issn0271-3683
dc.identifier.urihttp://hdl.handle.net/10347/32634
dc.language.isoenges_ES
dc.publisherTaylor and Francis Groupes_ES
dc.relation.publisherversionhttps://doi.org/10.1080/02713683.2018.1461910es_ES
dc.rightsCreative Commons Attribution-NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited.es_ES
dc.rights.accessRightsopen accesses_ES
dc.rights.urihttp://creativecommons.org/licenses/by-nc-sa/4.0/
dc.subjectCentral corneal thicknesses_ES
dc.subjectScheimpflug cameraes_ES
dc.subjectAnterior segment optical coherence tomographyes_ES
dc.subjectInterchangeabilityes_ES
dc.subjectNon-contact tono-pachymetryes_ES
dc.subjectUltrasound pachymetryes_ES
dc.titleComparison of Central Corneal Thickness measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomographyes_ES
dc.typejournal articlees_ES
dc.type.hasVersionAMes_ES
dspace.entity.typePublication
relation.isAuthorOfPublication1420ecc5-a082-4e70-8130-44dfeeadeed5
relation.isAuthorOfPublication.latestForDiscovery1420ecc5-a082-4e70-8130-44dfeeadeed5

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2018 Comparison of central corneal thickness.pdf
Size:
940.1 KB
Format:
Adobe Portable Document Format
Description:

Collections